On-Site Calibrations and Transportation
The HMK15 calibrator is easily trasferred from one place to
another. Two O-rings seal the transit cover on the chamber.
The optional transit bag helps to transport the calibrator so
that chambers stay in the vertical position. Furthermore, the
bag can be used for housing the calibrator during
calibration.
0606-174
Figure 11 Optional Transit Bag for HMK15
When transferring the calibrator from one place to another,
follow the instructions below:
- Turn the protective sleeve on the thermometer and place
the thermometer in the holder.
- Replace chamber covers with transit covers. Press
chamber covers on vacant chamber holders for
transportation.
- During transportation, keep the calibrator so that the
chambers stay as upright as possible. This way, very little
salt solution gets on transit covers and cleaning is easier.
The closer the transportation temperature is to the
temperature of the calibration site, the shorter the time
required for temperature stabilization. If the