Technical Reference _________________________________________________________________
16 ________________________________________________________________ M211638EN-A
Get_Parameter_Info
The Get_Parameter_Info command (ID 83h) reads the properties of a
single parameter from the HMM105. The command is useful for adapting
to parameters that have been added in a new software version. If the
requested parameter ID is not valid, data type Unknown Parameter ID
is returned.
Table 22 Get_Parameter_Info Invoke Message
Message Segment
Length
Content
I
2
C address
1 byte
I
2
C address (write)
Command ID
1 byte
Always 83h
Device address
1 byte
See Table 5 on page 4
Frame length
1 byte
Always 06h.
Data
1 byte
Parameter ID
Checksum
2 bytes
See section Checksum on page 9
Table 23 Get_Parameter_Info Response Message
Message Segment
Length
Content
I
2
C address
1 byte
I
2
C address (read)
Status
1 byte
See section Status Byte on page 9.
Command ID
1 byte
Always 83h
Device address
1 byte
See Table 5 on page 4
Frame length
1 byte
Always 12h.
Data
1 byte
Parameter ID
1 byte
Parameter datatype. See Table 24
below.
1 byte
Parameter length in bytes.
1 byte
Parameter persistence. Describes if the
parameter value is volatile (lost at reset)
or non-volatile (survives reset). See
Table 25 below.
8 bytes
Parameter name. If the name is shorter
than 8 bytes, extra bytes are filled with
00h.
Checksum
2 bytes
See section Checksum on page 9
Table 24 Parameter Data Types
Code
Meaning
0
Unknown parameter ID
1
Byte
2
Integer (16-bit)
3
Unsigned Integer (16bit)
4
Float (32-bit)
5
String
Table 25 Parameter Persistence
Code
Meaning
0
Void
1
Volatile
2
Non volatile
Vue de la page 17