And switch the station off for 30 s. Change
the HyperTerminal to the new baud rate
(Call/Disconnect,
File/Properties/Configure and Call/Call
again).
Configuring the composite message
The composite message is configured to
contain the desired fields only.
Which wind fields are included in the
composite message, are configured with
command like:
0WU,R=&01001100,U=M
The 0's and 1's disable and enable fields.
The first 0 or 1 defines if Dn (Wind
direction minimum) is to be sent, the next
defines Dm (Wind direction average) etc.
0=not included and 1=included.
The "bits" have the meanings:
Bit Field
1 Dn Wind direction minimum
2 Dm Wind direction average
3 Dx Wind direction maximum
4 Sn Speed minimum
5 Sm Speed average
6 Sx Speed maximum
7 spare
8 spare
U=M defines that metric units are used
(m/s). This application note will not repeat
every detail about these settings – refer to
the Vaisala manual.
The pressure, temperature and humidity
fields are configured with similarly:
0TU,R=&11010000,P=H,T=C
Bit Field
1 Pa Air pressure
2 Ta Air temperature
3 Tp Internal temperature
4 Ua Air humidity
5 spare
6 spare
7 spare
Bit Field
8 spare
P=H selects pressure in Hectopascals.
T=C selects temperature in Celsius.
The rain fields are configured this way:
0RU,R=&00100100,U=M,S=M
Bit Field
1 Rc Rain amount
2 Rd Rain duration
3 Ri Rain intensity
4 Hc Hail amount
5 Hd Hail duration
6 Hi Hail intensity
7 spare
8 spare
Finally any supervisor fields are switched
off:
0SU,R=&00000000
Now, test that the composite message
contains the correct fields: Send
command
0R0
Vaisala should respond something like
this:
0R0,Dm=091D,Sm=0.1M,Ta=21.6C,Ua=9.9P,Pa=1025…
Numbering the parameters
Number the order of the parameters:
0R0,Dm=091D,Sm=0.1M,Ta=21.6C,Ua=9.9P,Pa=1025…
1...2.......3.......4........5.......6.....
This is necessary later when configuring
the 7470.
Finally enable the automatic
transmission
Send command
0XU,I=1
Now the weather station should send only
one type of message beginning with 0R0.
Vaisala is done. Close or disconnect the
HyperTerminal.