HMP140A SERIES
Operating Manual U175en-1.5
6 1999-12-08
3.3 Electrical connections
The wiring can be done in either of the two ways shown in Figure 3.3.
mounting on a juction box surface mount
Vs
RHout
SHIELD
Tout
GND
Figure 3.3 Electrical connections
Outputs
0...20 mA (4...20 mA) 20...35 VDC 17...24 VAC
min.
50
)(
12
ohmR
L
+ VDC
0...1 V 9...35 VDC 9...24 VAC
0...5 V 15...35 VDC 12...24 VAC
0...10 V 20...35 VDC 16...24 VAC
Current consumption 6...10 mA + output currents
Electrical connections screw terminals for 0.5 - 1.5 mm
2
wires (AWG
15...20), stranded wires recommended
Commentaires sur ces manuels