User's Guide ______________________________________________________________________
64 __________________________________________________________________ M211280EN-A
10. Choose 1-point adjustment from the MI70 adjustment menu and
press SELECT.
0505-185
Figure 18 Example of the MI70 Adjustment Menu
11. Press READY when the reading has stabilized in the reference
condition (may take 30 minutes or more). You can follow the
stabilization from the GRAPH display.
0505-186
Figure 19 Example of the MI70 Adjustment Mode Graph
Display
12. Enter the correct reference value with the arrow buttons. Press
OK.
13. Confirm the adjustment by pressing YES.
14. The adjustment is done. Press BACK and EXIT to return to the
basic display.
15. Turn off the MI70 and detach the connection cable.