Appendix A _________________________________Probe Installation Kits and Installation Examples
VAISALA ______________________________________________________________________ 165
Pressure Tight Swagelok Installation Kits (for
HMT337)
RH Probe Installation
Swagelok installation kit for the relative humidity probe includes
Swagelok connector with ISO3/8" or NPT1/2" thread. Vaisala order
codes: SWG12ISO38 or SWG12NPT12.
0508-032
Figure 87 Swagelok Installation Kit for RH-probe
The following numbers refer to Figure 87:
1 = Relative humidity probe
2 = Duct connector
3 = ISO3/8" or NPT1/2" thread
4 = Swagelok connector
5 = Ferrules
Temperature Probe Installation
Swagelok installation kit for T-probe includes Swagelok connector with
either ISO1/8" or NPT1/8" thread. Vaisala order codes: SWG6ISO18 or
SWG6NPT18.
135 mm
33 mm
18 mm min
max 30 mm
12 3
4
P = max 10 bar
T = max 180 ºC
0508-016
Figure 88 Swagelok Installation Kit for T-Probe
The following numbers refer to Figure 88:
1 = T-probe
2 = Duct connector
3 = Swagelok connector
4 = Ferrules
Commentaires sur ces manuels