47
Technical Data
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Accuracy
Temperature range +10 ... +30 °C (+50 ... +86 °F)
0 ... 90 %RH ±2 %RH
90 … 100 %RH ±3 %RH
Temperature range -20 ... +10° C, +30 ... +60 °C
(-4 ... +50 °F, +86 ... +140 °F)
0 … 90 %RH ±3 %RH
90 … 100 %RH ±4 %RH
Temperature range -40 ... -20 °C (-40 ... -4 °F)
0 ... 100%RH ±4 %RH
Stability in typical HVAC applications ±0.5 %RH/year
Humidity sensor Vaisala HUMICAP
®
180R
TEMPERATURE
Measurement range -40 ... +60 °C (-40 ... +140 °F)
Accuracy
At +20 °C (+68 °F) ±0.3 °C (±0.54 °F)
Temperature dependence ±0.01 °C/ °C
Temperature sensor Pt1000 RTD Class F0.1 IEC 60751
CALCULATED PARAMETERS
Measurement range for dew
point temperature and wet bulb
temperature -40 ... +60 °C (-40 ... +140 °F)
Measurement range for enthalpy -40 ... 460 kJ/kg
(-10 ... +190 BTU/lb)
Accuracy of the calculated parameters should be
calculated at the actual condition based on the
RH and temperature specification. Accuracy at
20°C (68°F) and 80%RH:
Dew point ±0.7°C (1.2°F)
Wet bulb temperature ±0.5°C (0.9°F)
Enthalpy ±1.6kJ/kg (0.7 BTU/lb
Models
MODEL NUMBER TYPE OUTPUT SPECIAL FEATURES INGRESS PROTECTION
HMS110 Outdoor, RH+T 2-wire, current
output
Delivered with customer specific output
settings, including calculated humidity
parameters and special scaling of outputs.
IP65
HMS112 Outdoor, RH+T 2-wire, current
output
IP65
Operating Environment
Operating temperature range -40 ... +60 °C (-40 ... +140 °F)
Operating humidity range 0 ... 100 %RH
Maximum wind/flow speed 30 m/s
Storage temperature -40 ... +60 °C (-40 ... +140 °F)
Electromagnetic compliance EN61326-1, Industrial
Environment
Mechanics
Max wire size 1.5 mm
2
(AWG 16)
Standard housing color White (RAL9003)
Housing material PC + 10%GF (UL-V0 approved)
Inputs and Outputs
Analog output 4 ... 20 mA, loop powered
Loop resistance 0 ... 600 Ω
Supply voltage 20 ... 28VDC at 600 Ω load
10 ... 28VDC at 0 Ω load
Data input for
RDP100 Remote Panel Display
RS485,
Vaisala proprietary protocol
Spare Parts and Accessories
Remote Panel Display RDP100
Conduit fitting + O-ring (M16x1.5 / NPT1/2 Inch) 210675SP
Conduit fitting + O-ring (M16x1.5 / PG9, RE-MS) 210674SP
Fastening set HMS110 237805
Membrane Filter ASM210856SP
Terminal Block, Blue 236620SP
USB cable for PC connection 219690
Connection cable for HM70 hand-held meter 219980SP
HUMICAP
®
180R sensor HUMICAP180R
Ref. B211354EN-A ©Vaisala 2013
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Please contact us at
www.vaisala.com/requestinfo
www.vaisala.com
Scan the code for
more information
Humidity