USER'S GUIDE____________________________________________________________________
Selecting and Scaling the Analog
Output
0605-093
Figure 16 Analog Output Menus
To get analog measurement data, you need the analog output signal
cable, see section Options and Accessories Available on page 62. One
voltage signal channel 0...1.0V can be scaled for the selected quantity.
1. Connect the analog output signal cable connector to the
indicator base connector. Connect the screw terminal block as
follows:
- Brown wire: Common wire (-)
- Yellow-green wire: Signal (+)
2. Open the MENU. First press the right arrow key, and then press
the OPEN key.
3. Select Functions. Press the right arrow key.
4. Select Analog output, press the right arrow key.
5. Select Quantity and press the SET key to choose quantity.
6. Select quantity, press the SELECT key. (Only one quantity can
be chosen at a time. Only active quantities can be selected. To
change the active quantities, enter MENU, Display, Quantities
and units).