TECHNICAL REFERENCE___________________________________________________________
38 __________________________________________________________________ M211097EN-C
0212-040
Figure 12 Wind Direction Voltage Output with WS425 Cable and
Adapter Cable
Limitations for Output Signals
You can specify the minimum and maximum values for analog output
with the configuration parameters. The output is fixed to the specified
values, and the unit depends on the selected analog output mode.
Example
To limit the output 1 in voltage mode to a range of 0.1 … 5 V, set the
analog output minimum value to 0.1 and the analog output maximum
value to 5. Enter the following commands:
S aout1minv,0.1
S aout1maxv,5
For information on the parameters, refer to Vaisala WINDCAP
®
Ultrasonic Wind Sensor Series WMT700 User's Guide.