User's Guide _______________________________________________________________________
52 ___________________________________________________________________ M211024EN-C
1. Remove the filter from the probe by turning it counter-clockwise.
2. Remove the damaged sensor and insert a new one. Handle the new
sensor by the plastic socket. DO NOT TOUCH THE SENSOR
PLATE.
3. Perform a calibration and adjustment using the CRH command as
instructed in section Humidity Calibration (CRH) on page 58.
4. Attach a new filter on the probe. When using the stainless steel
filter, take care to tighten the filter properly (recommended force
5 Nm).
1209-033
Figure 16 Changing the Sensor
The following numbers refer to Figure 16 above:
1
=
The HUMICAP 180R sensor
2
=
Pt1000 sensor (do not touch)
Vue de la page 53