Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 95
Table 26 Other Commands
Output information about the device
Output information about the device in
POLL mode
Turn the serial interface echo ON/OFF
List present transmitter errors
All devices in POLL mode send their
addresses
List the currently available commands
Set the display backlight mode
Lock the menu or disable the keypad
View Modbus diagnostic counters
View information about connected modules
Set the value for pressure compensations
Limit highest value of RH output to 100%
Display the software version information
Set the value for pressure compensations,
temporarily
Getting Measurement Message from Serial Line
Starting Continuous Outputting
Use the R command to start the continuous output of measurements.
R<cr>
The default output format depends on the quantities your transmitter
supports. You can change the output message format and interval with
the FORM and INTV commands. See sections FORM on page 101 and
INTV on page 115.
Example:
>r
RH= 60.5 %RH T= 23.7 'C Tdf= 15.6 'C Td= 15.6 'C a= 13.0
g/m3 x= 11.1 g/kg Tw= 18.5 'C H2O= 17889 ppmV pw=
17.81 hPa pws= 29.43 hPa h= 52.3 kJ/kg dT= 8.1 'C
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars ‘*’.
Example:
>r
RH=***.* %RH T= 31.0 'C