Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 45
Sensor Functions
The active output version of HMP155 includes optional functions such as
chemical purge and sensor heating. These functions are described in
more detail in this section.
Chemical Purge (Optional)
In some specific applications the sensor gain may decrease gradually due
to an interference caused by a particular chemical present in the
measured gas, for example. The decrease of sensor gain due to an
interfering chemical and the effect of the chemical purge process are
illustrated below, see Figure 11 below. The sensor polymer absorbs the
inte
rfering c
hemical; and this reduces the ability of the polymer to absorb
water molecules and consequently the sensor gain decreases. In chemical
purge, heating the humidity sensor to a temperature level of
approximately +180 °C for several minutes evaporates the interfering
chemical.
The purge function starts with heating stage, continues with settling and
when the temperature of the sensor is decreased, the probe returns to
normal mode. The whole cycle takes about 6 minutes.
NOTE
Chemical purge function locks the output values for about 6 minutes.
NOTE
Avoid using the purge function in low (below freezing) temperatures.
Low temperatures decrease the effectivity of purge. Also, the lower the
temperature, the longer it takes for the sensor to return to normal
functionality.