Model HMP45C Temperature and Relative Humidity Probe
Because of the inverse relationship between relative humidity and air
temperature, finding the mean relative humidity is meaningless. A more useful
quantity is the mean vapor pressure. The mean vapor pressure can be
computed on-line by the datalogger as shown in the following examples.
TABLE 7-1. Wiring for Vapor Pressure Examples
Color Description CR1000 CR10(X)
Yellow Temperature SE 2 (1L) SE 3 (2H)
Blue Relative Humidity SE 1 (1H) SE 4 (2L)
White Signal Reference
AG
Jumper from
SW12V Control
C1
Red Power SW12V SW12 V
Black Power Ground
AG
Clear Shield
G
CR1000 Program that Computes Vapor Pressure and Saturation Vapor Pressure
'CR1000 program that calculates Vapor Pressure
Public AirTC
Public RH
Public RH_Frac, e_Sat, e_kPa
DataTable(Temp_RH,True,-1)
DataInterval(0,60,Min,0)
Average(1,AirTC,IEEE4,0)
Sample(1,RH,IEEE4)
Sample(1,e_kPa,IEEE4)
EndTable
BeginProg
Scan(1,Sec,1,0)
'HMP45C Temperature & Relative Humidity Sensor measurements AirTC and RH:
SW12 (1 )
Delay(0,150,mSec)
VoltSE(AirTC,1,mV2500,2,0,0,_60Hz,0.1,-40.0)
VoltSE(RH,1,mV2500,1,0,0,_60Hz,0.1,0)
SW12 (0)
If RH>100 And RH<108 Then RH=100
'Calculate Vapor Pressure
'Convert RH percent to RH Fraction
RH_Frac = RH * 0.01
'Calculate Saturation Vapor Pressure
13