USER'S GUIDE____________________________________________________________________
60 __________________________________________________________________ M211399EN-B
Repair Maintenance
Repair services are available through Vaisala Service Centers. For
contact information of Vaisala Service Centers, see
www.vaisala.com/servicecenters.
Replacing the HTM10 Module
If you cannot restore the measurement accuracy of the transmitter by
calibration and adjustment, you can replace the HTM10 module inside
the transmitter. HTM10 module is the small separate component board
with the HUMICAP® sensor; see Figure 4 on page 15.
NOTE
User calibration settings (adjustment by trimmers or service port) are
stored in the HTM10 module. If you replace the module, you do not need
to undo the previously applied correction.
CAUTION
Handle the HTM10 module carefully. When reinstalling the transmitter
body to the mounting base, avoid touching the HUMICAP® sensor or
the module.
1. Disconnect the transmitter body from the mounting base.
2. With your fingers, push apart the two plastic holders that hold the
HTM10 module and pull out the module.
1112-028
Figure 22 Replacing the HTM10 Module