User's Guide _______________________________________________________________________
84 ___________________________________________________________________ M211501EN-C
Analog Output Error State
GMP231 sets the analog output channel into a defined error level instead
of the measured result in two situations:
- Probe detects a measurement malfunction. This means an actual
measurement problem, such as sensor damage or unsuitable
environmental conditions.
- Measured value(s) are well outside the scaled output range. For a
more detailed explanation, see section Analog Output Overrange
Behavior on page 19.
The default error level depends on the output type:
- For 4 ... 20 mA output, the default error level is 3.6 mA
- For 0 ... 20 mA output, the default error level is 23 mA.
The probe resumes normal operation of the analog output when the cause
of the error state is removed.
Technical Support
For technical questions, contact the Vaisala technical support by e-mail at
helpdesk@vaisala.com. Provide at least the following supporting
information:
- Name and model of the product in question
- Serial number of the product
- Name and location of the installation site
- Name and contact information of a technically competent person who
can provide further information on the problem.
Product Returns
WARNING
When returning a product for calibration or repair, make sure it has not
been exposed to dangerous contamination, and is safe to handle without
special precautions.
If the product must be returned for service, see www.vaisala.com/returns.
For contact information of Vaisala Service Centers, see
www.vaisala.com/servicecenters.