Vaisala GMP343 Manuel d'utilisateur Page 57

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 94
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 56
Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 55
Setting the Analog Output Overrange Clip
This command clips the analog output signal at the high value when the
signal goes over range. In case of the current output the current will not go
higher than 20 mA or below the low value current (set with ILOW). The
voltage output will not go higher than the voltage set with UHIGH.
ACUT x <cr>
SAVE <cr>
x=ON/OFF
>acut
ACUT : OFF ? ON
>acut
ACUT : ON ? OFF
>
>save
EEPROM saved successfully.
>
Testing the Analog Output
When you want to test the operation of the analog output use this command
to force the output to the set value. The value in the analog output can then
be measured with a current/voltage meter. The set value remains valid until
you give the ATEST command without a value or RESET the transmitter.
ATEST <cr>
x=the given test value (mA or V)
>atest 1
Test voltage set at 1 V. Use command without any parameters
to stop test mode.
>atest
Voltage test mode stopped.