User’s Guide ______________________________________________________________________
58 ___________________________________________________________________M210514EN-E
Linear and Multipoint Corrections
Shows the last linear and multipoint correction values.
CORR <cr>
>corr
LC : OFF
Linear correction
Reading Reference
0.00 0.000
1000.00 1000.000
MPC : OFF
Multipoint correction
Reading Reference
0.00 0.000
1000.00 1000.000
>
Error Messages
The ERRS command shows the error messages received. For a list of
possible errors, refer to Table 7 on page 72.
ERRS <cr>
>errs
No errors detected.
>errs
ERROR E02: IR source failure.
ERROR E06: Temperature measurement failure (recovered 1 h 9
min ago).
WARNING W01: Watchdog reset.
>
The ERRS R command resets the error states.
ERRS R <cr>
>errs r
OK: No errors detected.
Error states cleared.