Vaisala Calibration of Digital Transmitters with HMI41 Manuel d'utilisateur Page 20

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 36
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 19
User's Guide ______________________________________________________________________
18 __________________________________________________________________ M210777EN-B
In this case, perform the calibration again. Whether the calibration
was successful or not, HMI41 always returns to show the selected
calibration mode:
RH
or
RH
After having successfully completed the calibration, turn HMI41 off
and disconnect the cable.
Two-Point Calibration
In two-point calibration mode, both offset and gain corrections are
made. This calibration mode is more accurate than offset or gain
calibrations. For performing a two-point calibration you need two
reference points with a difference of at least 50 %RH between them.
NOTE
Two-point calibration is not supported for HMT100.
Select two-point calibration with buttons and :
RH
Press ENTER and the following text appears on the display for about
1 ... 2 seconds:
RH
Vue de la page 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 35 36

Commentaires sur ces manuels

Pas de commentaire