Chapter 5 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 37
Set Output Interval
Use the INTV command to change the output interval of the
automatically repeating measurement messages. The measurement
messages are repeated in the RUN mode, or after the R command has
been given.
INTV [n xxx]<cr>
where
n
=
time interval, range 0 ... 255
xxx
=
time unit = "S", "MIN", or "H"
The shortest output interval (with n = 0) outputs the measurement
messages as quickly as the transmitter’s internal measurement cycle
produces them, without additional delay.
Example:
intv 1 min
Output interval: 1 MIN
Output a Reading Once
Use the SEND command to output a single measurement message. The
output message is sent in the format that has been set using the FORM
command.
SEND [address]<cr>
where
address
=
Address of the transmitter, range 0 ... 255.
Specifying the address is only necessary if the target
transmitter is in POLL mode, and has not been accessed
with the OPEN command.
Example:
send
Tdf= 15.4 'C Tdfatm= 16.2 'C H2O= 18539 ppm P= 0.959 bara
Pnorm= 0.951 bara Rhoo= 1.1 kg/m3 T= 25.4 'C