
HMP240
Appendix 6: HMP243 wiring diagram M210300en
114
- the humidity measurement cannot be completed. Check the HUMICAP
sensor and humidity measurement wires and their solderings on the
component board (TP9,TP10 and TP11/15). If you cannot find any faults, send
the transmitter to Vaisala for repair.
E46 f (Ud2 ) out of range
- the measurement hybrid is probably damaged. Send the transmitter to
Vaisala for repair.
E47 f (Uk1 ) out of range
- the measurement hybrid is probably damaged. Send the transmitter to
Vaisala for repair.
E48 f (Uk2 ) out of range
- the measurement hybrid is probably damaged. Send the transmitter to
Vaisala for repair.
The LEDs display these error types as follows:
¡¡¤¤
no frequency at all
¡¡¤¡
a frequency missing from the RH channel
¡¡¡¤
a frequency missing from the T channel
The second error type indicates erroneous y-values (used in internal calcula-
tions):
Display: Serial bus:
E51 T y-value out of range
- the temperature measurement cannot be completed. Check the temperature
sensor when only a humidity probe is used, and both temperature sensors if
also an additional temperature probe is used. Check also the temperature
measurement wires and their solderings on the component board
Commentaires sur ces manuels