HMP243
M210300en Appendix 7: Re-gaining
125
5. SERIAL COMMANDS
1.2.1 REG Re-gaining settings
REG <cr>
REG nnn <cr>
nnn = ON or OFF
Command REG is used to check the re-gaining parameter settings and to
activate or inactivate the automatic re-gaining.
Examples: checking the current settings
>reg
REG OFF automatic re-gaining is inactive
Interval min : 720
Duration s : 120
Temp diff : 0.50
Settl time s : 240
k % : 95.00
>
activating the automatic re-gaining
>reg on
REG ON
>
deactivating the automatic re-gaining
>reg off
REG OFF
>
1.2.2 CR Re-gaining parameters
CR <cr>
Command CR can be used to check the values of re-gaining parameters.
Example:
>cr
Interval min : 720
Duration s : 120
Temp diff : 0.50
Settl time s : 240
k % : 95.00
>