USER'S GUIDE_______________________________________________________________________
18 ___________________________________________________________________ M010022EN-C
CHAPTER 7
TECHNICAL DATA
Technical specifications
Measuring range
GMT221 0...2% CO
2
, 0...3% CO
2
,
0...5% CO
2
, 0...10% CO
2
,
0...20% CO
2
GMT222 0...2000 ppm, 0...3000 ppm,
0...5000 ppm, 0...7000 ppm,
0...10 000 ppm
Accuracy at 25°C against certified factory references
GMT221 <±[0.02% CO
2
+ 2% of reading]
GMT222 <±[20 ppm CO
2
+ 2% of reading]
(including repeatability and calibration uncertainty)
Non-linearity <±0.5 %FS
Temperature dependence
(typically) -0.1 %FS / °C (When temperature
rises, the output decreases, default
25 °C (77 °F))
Pressure dependence (typically) +0.15 % of reading/hPa (When
pressure rises, the output increases,
default 1013 hPa (1 atm))
Long-term stability <
±5 %FS / 2 years
Response time (63%) 20 seconds (GMT221)
30 seconds (GMT222)
Operating conditions
Operating temperature range -20…+60 °C
Humidity range 0 - 100 %RH (non-condensing)
General
Analog outputs 0 - 20 mA or 4 - 20 mA and
0 - 10 V
Resolution of analog outputs 0.03 %FS