Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 97
Table 28 Selection of Output Modes
Mode Output Commands used
STOP Only by command All (default mode)
RUN Automatic output Only command S
POLL Only with command SEND [addr] Use with RS-485 buses, see
RS-422/485 Interface on page
53.
SEND One message at power-up
Selected output mode will be activated after power outages.
INTV
Use the command INTV[xxx yyy] to set the outputting interval for the
RUN mode.
Syntax: INTV[xxx yyy]
where
xxx
= Output interval (0 ... 255). 0: the fastest possible output rate.
y
yy
= Unit (s, min or h)
Example:
>INTV 10 min
Output intrv. : 10 min
>
ECHO
Use the command ECHO [x] to set the user port echo. The command
either enables or disables echo of characters received.
Syntax: ECHO [x]
where
x = ON (default) or
= OFF
NOTE
You can use the SERI, SMODE, INTV and ECHO commands to
change/view the user port settings even if you are currently connected to
the service port.