User's Guide ______________________________________________________________________
34 __________________________________________________________________ M211060EN-C
View Software Version
Use the VERS command to display the software version of the probe.
VERS<cr>
Example:
vers
HMP60 / 1.00.0
Serial Line Output Commands
Start Measurement Output
Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line.
For HMP60 and HMP110, the output always includes readings for
temperature, RH and Td. For HMP110T, the output includes only
temperature.
R<cr>
Example (HMP60 and HMP110):
r
T= 22.6 'C RH= 22.8 %RH Td= 0.3 'C
T= 22.6 'C RH= 22.5 %RH Td= 0.2 'C
T= 22.6 'C RH= 22.5 %RH Td= 0.2 'C
Example (HMP110T):
r
T= 22.6 'C
T= 22.6 'C
T= 22.6 'C
Outputting the results continues in intervals issued with the command
INTV. You can stop the output by entering the S command.
Stop Measurement Output
Use the S command to stop the continuous measurement output.
S<cr>