User’s Guide ______________________________________________________________________
52 ___________________________________________________________________ M010056EN-J
Example:
>asel rh t
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo: -40.00 ’C
Ch2 T hi: 100.00 ’C
>
S Scaling Analog Outputs
Syntax: S [zz aa.a bb.b]<cr>
Turn the internal DIP switch Calibration Enabled/Disabled to position
ON before the scaling the analog outputs and return it to position OFF
after scaling.
Example:
>srh 0 100
RH lo: 0.00 %RH
RH hi: 100.00 %RH
>
where
zz = Quantity (RH, T, Td, x, a, Tw, ppm, aw, RS, Ts)
aa.a = Lower limit of the quantity
bb.b = Upper limit of the quantity
Vue de la page 53