Chapter 12 ______________________________ Calibration and Adjustment of HM70 Series Probes
VAISALA _______________________________________________________________________ 65
2-Point Adjustment
Follow the steps from 1 to 4 and continue as follows:
5. Select 2-point adjustment. Press SELECT to continue.
6. Set the probe to a lower reference temperature. You can follow
the stabilization from the GRAPH display.
7. Press READY when the reading is stabilized. Give the lower
reference temperature by using the arrow buttons. Press OK.
8. Set the probe to the higher reference temperature. You can
follow the stabilization from the GRAPH display. Press
READY when the reading is stabilized.
9. Give the higher reference temperature by using the arrow
buttons. Press OK.
10. Confirm the adjustment, press YES. (By pressing NO you
return to adjustment mode display and no changes are made). If
the difference between two references is less than 30 ºC,
adjustment can not be done.
11. Calibration is carried out. Press BACK to exit the adjustment
mode and EXIT to return to the basic display.
Last Adjustment Date
To see the last adjustment date, select Last adjustment date. You can
check this date also from the
Device information, see page 27.