Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 31
FTIME, FDATE Adding time and date to R and SEND outputs
FTIME x
FDATE x
x = ON or OFF
Command will enable/disable output of time and date to the serial
line.
>send
aw= 0.2 aw T= 31.0 'C
>ftime on
Form. time : ON
>send
03:47:59 aw= 0.2 aw T= 31.0 'C
>fdate on
Form. date : ON
>send
2000-01-01 03:48:03 aw= 0.2 aw T= 31.0 'C
>
UNIT Selecting metric or non-metric output units
UNIT x
x = M or N
M = metric units
N = non-metric units
Quantity Metric Unit Non Metric Unit
aw - -
T
°C °F
ppm
*
ppm ppm
*
for transformer oil only