USER'S GUIDE____________________________________________________________________
32 __________________________________________________________________ M210474EN-B
Setting, scaling and testing the analog outputs
You can select for the two outputs
- current range (0...20 mA/4...20 mA) and
- output parameter (aw/ ppm
*
/ T).
and scale the two outputs according to your needs.
*
for transformer oil only
AMODE Setting the analog outputs (0...20 mA/4...20 mA)
AMODE ch1 ch2
where ch1 and ch2: I0 = 0...20 mA
I1 = 4... 20 mA
>amode i1 i1
Ch1 output mode: 4...20mA
Ch2 output mode: 4...20mA
>
If the output includes a reminder 'remember to set jumpers', please
ignore the reminder.
ASEL Selecting the parameter for the analog outputs
ASEL xxx yyy
where: xxx = quantity of channel 1 and yyy = quantity of channel 2.
Use abbreviations shown in the table below.
Quantity abbreviation
aw water activity aw
T Temperature T
ppm (for transformer oil only) ppm
Examples:
>asel aw t
Ch1 aw lo : 0.00 ?