Vaisala HUMICAP MMT330 Guide de l'utilisateur Page 120

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 181
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 119
User's Guide _______________________________________________________________________
118 __________________________________________________________________ M210733EN-G
AMODE/ASEL
Use the serial line to select and scale the analog output quantities. Check
the analog output modes with the AMODE command.
AMODE<cr>
Example:
>amode
Ch1 output : 0...1V
Ch2 output : 0...1V
>
Select and scale the quantities for the analog outputs with the command
ASEL. Note that the optional quantities can be selected only if they have
been selected when ordering the device.
ASEL [xxx yyy zzz]<cr>
where
xxx
=
Quantity of channel 1
yyy
=
Quantity of channel 2
zzz
=
Quantity of the optional analog output channel 3
Always enter the quantities for all outputs. The MMT330 measures the
following quantities:
- Water activity (aw)
- Temperature (T) (metric unit: ºC, non-metric unit:ºF)
- ppm for transformer oil only (H
2
O) (optional quantity)
Use the ASEL [xxx yyy] command as shown in the example below when
using a device with two analog outputs.
Example:
>asel aw t
Ch1 aw low : 0.00
Ch1 aw high : 1.00
Ch2 T low : -40.00 'C
Ch2 T high : 60.00 'C
>
Vue de la page 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 180 181

Commentaires sur ces manuels

Pas de commentaire