User's Guide _______________________________________________________________________
146 __________________________________________________________________ M210733EN-G
Table 30 Error Messages
Humidity sensor measurement
malfunction.
Check the integrity of the humidity probe
and the probe cable. Clean the probe from
dirt, water, ice or other contaminants.
Humidity sensor short circuit
Humidity sensor open circuit
Check the integrity of the humidity probe
and the probe cable.
Temperature sensor open
circuit.
Temperature sensor short
circuit.
Check the integrity of the humidity probe
and the probe cable. Clean the probe from
dirt water, ice or other contaminants.
Temperature measurement
malfunction
Temperature sensor current
leak.
Check the integrity of the humidity probe
and the probe cables. Clean the probes
from dirt, water, ice or other contaminants.
Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Checksum error in the internal
configuration memory
Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.
Internal EEPROM read error
Internal EEPROM write error
Add-on module 1 (or 2)
connection failure
Turn off the power and check the module
connection. Turn on the power.
Device internal temperature
out of range
Ensure that the operating temperature is
within the valid range.
software checksum error
Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Internal ADC reference voltage
out of range
Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.
Internal analog output
reference voltage out of range
Configuration switches for
analog output 1/2/3 set
Check and re-set the switches, see Figure
58 on page 116 and Figure 30 on page 53.
Internal error in add-on module
1 (or 2)
Disconnect the power and check the
module connection.
installed in incorrect add-on
Disconnect the power and change the
communication module to another module
Unknown/incompatible module
installed in add-on module slot
Ensure that the module is compatible with
the MMT330.
Internal analog voltage out of
range
Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Internal system voltage out of
range
Check that the supply voltage is correct
and the power supply can provide enough
power for the instrument.
Commentaires sur ces manuels