Chapter 4 ________________________________________________________________ Operating
VAISALA _______________________________________________________________________ 19
.CALD Setting the Calibration Date
With this command, the last calibration date can be stored in the
memory of the barometer.
.CALD.xxx
xxx = calibration date; max. 15 characters
Example:
.CALD.991015<cr>
Setting the RS485 Terminating
Resistor On/Off
The PTB210 series barometers have a built-in RS485 terminating
resistor, which is set, as default, OFF. The resistor can be set either ON
or OFF with the .ROFF/.ROFF commands. If the barometer is in the end
of the serial line, the resistor should be set ON, especially, when long
cables are used.
In case several barometers are connected to same line, only the
resistor of the last one is set ON, with the others it should be OFF. This
prevents formation of excessive load on the line. The maximum
amount of barometers on same line is 32.
At the end of the serial bus there must be a PTB210 barometer, a
dynamic line adapter or a line master (see figure below).
HOST
COM PUTER
PTB210
R
X/T
X
HI
LO
RX / T X
HI
LO
RX / T X
HI
LO
RX / T X
HI
LO
120R
33nF
DYNAM IC LINE TERMINATION REQUIRED
IF NO DEVICE AT THE END OF THE BUS
..........
..........