Vaisala PTB220 Series Guide de l'utilisateur Page 74

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 113
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 73
USER'S GUIDE_______________________________________________________________________
68 ___________________________________________________________________ M210194EN-A
>errs <cr>
E00 Nothing special to report
>
Complete list of error codes:
E00 Nothing special to report
E10 CPU EEPROM ackn. error
E20 CPU EEPROM csum error
E11 P1 EEPROM ackn. error
E14 DAC EEPROM ackn. error
E21 P1 EEPROM csum error
E24 DAC EEPROM csum error
E31 P1 serial number error
E41 P1 f out of range
E51 P1 y-value out of range
E61 P1 pressure out of range
E71 P1 P difference too large
E12 P2 EEPROM ackn. error
E22 P2 EEPROM csum error
E32 P2 serial number error
E42 P2 f out of range
E52 P2 y-value out of range
E62 P2 pressure out of range
E72 P2 P difference too large
E13 P3 EEPROM ackn. error
E23 P3 EEPROM csum error
E33 P3 serial number error
E43 P3 f out of range
E53 P3 y-value out of range
E63 P3 pressure out of range
E73 P3 P difference too large
TEST n Testing transducer operation
TEST n <cr>
where n = number of the pressure transducer (1, 2 or 3)
The TEST n command is used to output various measured quantities
and calculated values that characterize the measurement performed by
a transducer.
For transducer P1, the listing looks as follows:
>test 1 <cr>
204736 254733 321328 246633 231962
13.8283 11.1142 8.8108 11.4792 12.2053
0.57134 0.76660 -0.74368 0.04356 26.1
1033.553 1033.313 1033.313 1033.313 1033.313
1033.313 1033.313 -0.010 -0.007 0.010
>
For transducers P2 and P3, the listings can be obtained in a similar
way.
Vue de la page 73
1 2 ... 69 70 71 72 73 74 75 76 77 78 79 ... 112 113

Commentaires sur ces manuels

Pas de commentaire