User's Guide ______________________________________________________________________
74 ___________________________________________________________________M210983EN-A
Use the UNIT command to change the measurement unit for a quantity.
Specify the quantity and the desired unit for it.
Example:
>unit P mmhg
P : mmHg
P3h : Pa
P1 : Pa
P2 : Pa
DP12 : Pa
HCP : Pa
QFE : Pa
QNH : Pa
>
Use the UNIT ?? command to list the available measurement units for
the quantities.
User Port Serial Settings
SERI
Use the serial line command SERI [b p d s] to set communication
settings for the user port.
CAUTION
Changing the User Port serial settings may prevent the PTB330
barometer from operating with the MI70 indicator. To restore the
default settings, connect to the serial line using the service port, and
give the following commands:
SERI 19200 N 8 1
SMODE TS
ADDR 0
Vue de la page 75