Vaisala PTU300 Guide de l'utilisateur Page 112

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 217
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 111
User's Guide _______________________________________________________________________
110 __________________________________________________________________ M210796EN-G
Using Serial Line
PRES and XPRES
Command XPRES must be used if the value is changed frequently (for
example, by an automatic system that updates the value). Its value is not
retained at reset, and when set to 0, last value set with PRES is used
instead. Use the serial line and do the following:
PRES [aaaa.a]<cr>
XPRES [aaaa.a]<cr>
where
aaaa.a
=
Absolute process pressure (hPa)
Example:
>pres
Pressure : 1013.00 hPa ?
>pres 1010
Pressure : 1010.00 hPa
>
Table 31 Conversion Factors for Pressure Units
From
To: hPa
mbar
1
Pa N/m2
0.01
mmHg torr
1.333224
inHg
33.86388
mmH
2
O
0.09806650
inH
2
O
2.490889
atm
1013.25
at
980.665
bar
1000
psia
1)
68.94757
1) psia = psi absolute.
Example:
29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa
PFIX
Use the PFIX command to select either P inputted or P measured.
- When PFIX is On, fixed PRES value is used
- When PFIX is Off, measured PRES value is used
Vue de la page 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 216 217

Commentaires sur ces manuels

Pas de commentaire