Vaisala Drycap DMT152 Guide de l'utilisateur Page 55

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 70
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 54
Chapter 5 ________________________________________________________ Serial Line Operation
VAISALA ________________________________________________________________________ 53
Other Commands
Opening the Transmitter in POLL Mode
Use the OPEN command to connect to a transmitter that is in POLL
mode.
OPEN [aa]<cr>
where
aa
=
address (0 ... 99)
Example:
open 1
DMT152 1 line opened for operator commands
Closing the Connection to a Transmitter
in POLL Mode
The CLOSE command closes the connection to a transmitter.
CLOSE<cr>
Example:
close
line closed
Display Command List
HELP<cr>
Set Time
Use the TIME command to view and set the transmitter clock. The clock
does not have a battery backup – it is reset to 00:00:00 when the
transmitter is powered up or reset.
TIME [hh mm ss]<cr>
where
hh
=
Hours (24-hour clock)
mm
=
Minutes
ss
=
Seconds
Vue de la page 54