User's Guide _______________________________________________________________________
46 ___________________________________________________________________ M210474EN-C
Scale Analog Outputs
Syntax: ASCL<cr>
Example:
>ascl
Ch1 T lo : -40.00 'C ? -20
Ch1 T hi : 60.00 'C ? 40
Ch2 ppm lo : 0.00 ppm ? 0
Ch2 ppm hi : 5000.00 ppm ? 3000
>ascl
Ch1 T lo : -20.00 'C ?
Ch1 T hi : 40.00 'C ?
Ch2 ppm lo : 0.00 ppm ?
Ch2 ppm hi : 3000.00 ppm ?
The ppm calculation is only used for transformer oils.
Test Analog Outputs
Syntax: ITEST aa.aaa bb.bbb<cr>
The operation of the analog outputs is tested by forcing the outputs to
given values. The values in the analog outputs can then be measured with
a current/voltage meter.
Where
aa.aaa
=
Current/voltage value to be set for channel 1
bb.bbb
=
Current/voltage value to be set for channel 2
Example:
>itest 8.30 6.40
8.30000mA 2c30 6.40000mA 2278
>
The ITEST command output includes internal debugging values shown
in hexadecimal format (2c30 and 2278 in the example above).
The set values remain valid until you issue the ITEST command without
readings or reset the transmitter.