The port connections are as follows:
pin 1: ch+ (+VDC supply voltage)
pin 2: ch- (-VDC supply voltage)
pin 3: test (for service use only)
pin 4: shield
When using 12V loop voltage, the resistance must be less than 90
ohms. An approximate can be calculated by following formula:
Rmax = (V
loop
- 10V)/20mA
You can scale the output to any output range. It is recommended to
have the scaling within the measuring range to get accurate
measurements.
1. Connect the analog output signal cable connector to the
connector, see Figure 21 on page 73.
2. Open MENU: press .
3. Select Interfaces, press .
4. Select Analog outputs, press .
5. Select Output 1, press .
6. Select Quantity, press SET , select the quantity for the analog
output and press SELECT.
7. Select Scale 4 mA, press SET.
8. Press CLEAR and give the lowest value (value represented by 4
mA current) by using numerical buttons. Press OK.
9. Select Scale ...20 mA, press SET.
10. Press
CLEAR and give the highest value (value represented by
20 mA current) by using numerical buttons. Press OK.
11. Select ENABLED WHEN, press SELECT. Now you can select the
measurement status that activates the analog outputs.
STOP: output is enabled when dewpoint measuring status is
STOP.
ERR: output is enabled when dewpoint measuring status is
ERR.
OK: output is enabled when dewpoint measuring status is OK