User's Guide ______________________________________________________________________
88 __________________________________________________________________ M210205EN-D
Example:
>intv 1 min
Output interval: 1 MIN
>
FORM Defining the Output Format
Syntax: FORM [amount of decimals] [quantity] [u]
[status] [date][time] [addr]<cr>
With the FORM command you can determine the output form of
commands R and SEND. The following output fields can be defined
after the command FORM:
The fields can be given in any order, except the unit field U, that can
be given only after a quantity or after an error limit value. You can
give only one field or several fields and write the field descriptions
either in capital or small letters. Leave a space between the fields. The
maximum amount of characters in the command is 80.
BASIC FIELDS
[amount of decimals]
Type number of digits before and after the decimal point.
As an example: giving 2.1 before the RH quantity outputs a
reading with two digits before the point and one after the
point, as follows: 80.1. You can define different amount of
decimals before every quantity, if needed.
Type 0.0 to return the default setting