Vaisala DRYCAP DMT143 Guide de l'utilisateur Page 54

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 76
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 53
USER'S GUIDE ____________________________________________________________________
52 ___________________________________________________________________ M211435EN-E
Set User Adjustment Parameters
Use the LI command to set the user adjustment parameters directly. This
command is useful for restoring the desired adjustments, and removing
the effects of user adjustment without having to perform the factory reset.
CAUTION
This command is provided primarily for applying an offset to the Tdf
measurement. Do not adjust the other parameters unless instructed to do
so by a Vaisala Service Center.
LI<cr>
After entering the command, you will be prompted to enter new values
one at a time. Press ENTER without giving a value to move to the next
parameter.
Example (shows a +0.8 °C Tdf offset being applied):
li
RH offset : 0.00000000E+00 ?
RH gain : 1.00000000E+00 ?
T offset : 0.00000000E+00 ?
T gain : 1.00000000E+00 ?
Tdf offset: 0.00000000E+00 ? 0.8
Tdf gain : 1.00000000E+00 ?
Vue de la page 53