Vaisala DRYCAP DMT345 Guide de l'utilisateur Page 118

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 185
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 117
User's Guide _______________________________________________________________________
118 __________________________________________________________________ M210762EN-G
Analog Output Settings
The analog outputs are set in the factory according to the order form. In
case you want to change the settings, follow these instructions.
Changing Output Mode and Range
Both output channels have their own dip switch module with 8 switches,
see the position in Figure 2 on page 23 (dip switches for analog output
settings).
1. Select the current/voltage output, switch ON either of the switches,
1 or 2.
2. Select the range, switch ON one of the switches from 3 to 7.
0503-045
Figure 60 Current/Voltage Switches of Output Modules
The following numbers refer to Figure 60 above:
1
=
Current/voltage selection output switches (from 1 to 2)
2
=
Current/voltage range selection switches (from 3 to 7) in
analog output 1 and 2
3
=
Switches for service use only. Keep in OFF position always.
NOTE
Only set one mode and one range switch to ON for one output channel.
The rest of the switches must remain OFF.
Vue de la page 117
1 2 ... 113 114 115 116 117 118 119 120 121 122 123 ... 184 185

Commentaires sur ces manuels

Pas de commentaire