Vaisala DRYCAP DMT345 Guide de l'utilisateur Page 154

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 185
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 153
User's Guide _______________________________________________________________________
154 __________________________________________________________________ M210762EN-G
11. When stabilized, type the high end reference value after the
question mark and press Enter.
>fcrh
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
RH : 75.45 Ref2 ? c
RH : 75.57 Ref2 ? c
RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>
12. The OK indicates that the adjustment has succeeded and the new
calibration coefficients are calculated and stored. Enter the
adjustment information (date and text) to the memory of the
transmitter, see section Adjustment Information on page 151.
13. Press the ADJ button on the motherboard to close the adjustment
mode.
14. Take the probe out of the reference conditions and replace the
filter.
Vue de la page 153
1 2 ... 149 150 151 152 153 154 155 156 157 158 159 ... 184 185

Commentaires sur ces manuels

Pas de commentaire