Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 119
UNDELETE
Similarly to the DELETE command, the UNDELETE command is used
without any arguments. It will recover all deleted data that has not been
overwritten yet.
Analog Output Settings
The analog outputs are set in the factory according to the order form. In
case you want to change the settings, follow these instructions. See
section Third Analog Output on page 54.
Changing Output Mode and Range
Both output channels have their own DIP switch module with 8 switches;
see the position in Figure 2 on page 22 (DIP switches for analog output
settings).
1. Select the current/voltage output; switch ON either of the switches,
1 or 2.
2. Select the range; switch ON one of the switches from 3 to 7.