Chapter 2 __________________________________________________________ Product Overview
VAISALA _______________________________________________________________________ 25
Warmed Probe HMT337
Temperature difference between the probe and external environment can
cause a risk of condensation on the sensor. A wet probe cannot observe
the actual humidity in the ambient air. If the condensed water is
contaminated, the life span of the probe may shorten and calibration may
change.
HMT337 probe shall be used in applications where condensation can
occur due to high humidity and rapid humidity changes. The warmed
probe is heated continuously so that its temperature is always higher than
in environment. This prevents condensation on the probe. The power
consumption of the warmed probe is slightly higher than other probes.