Configuring Measurement Parameters
Select Analog Output Parameter and
Scaling
Select and scale the parameters for the analog outputs with the ASEL
command. Note that the optional parameters can be selected only if they
have been selected when ordering the device.
ASEL [xxx yyy]<cr>
Always enter all the parameters for all outputs. The MMT162 measures
the following parameters:
- water activity (aw)
- temperature (T) (metric unit: ºC, non metric unit: ºF)
- ppm for transformer oil only (H
2
O)
Example:
asel
Ch0 aw lo : 0.00 ?
Ch0 aw hi : 1.00 ?
Ch1 T lo : -20.00 'C ?
Ch1 T hi : 80.00 'C ?
Set Analog Output Mode (mA/V)
The AMODE command can be used to change the analog output mode
of channels 1 and 2. When used without parameters it displays the
current settings and with parameters new settings can be set. The
available parameters are listed in Table 9 below.
Table 9 Analog Output Modes