9. The first adjustment point is now stored. Now you have to choose:
- Press Enter once to continue to the next step of the procedure
(two-point adjustment). The adjustment sequence continues
to the second reference point:
T : 22.3009 2. ref ?
- Press Enter twice to exit the procedure now, completing a
one-point adjustment only. The adjustment sequence ends
with the text OK. Continue to step 12.
OK
10. Transfer the probe to the second temperature reference. Do not
disconnect the cable from the transmitter or the adjustment
sequence will be interrupted.
11. Wait for the temperature to stabilize. Type c and press Enter to
update the reading and verify that the reading has stabilized:
T : 54.1243 2. ref ? c
T : 54.1442 2. ref ?
12. When the reading is stable, enter the reference temperature after the
question mark and press Enter:
T : 55.1841 2. ref ? 55.1
OK
13. The temperature adjustment is now complete. Enter the l
command to verify that the calibration coefficients for T have
changed.
14. Remove the MMT162 from the reference environment and reattach
the filter.
Analog Output Adjustment
In the analog output calibration the analog output is forced to the
following values:
- Current output: 2 mA and 18 mA.
- Voltage output: 10 % and 90 % of the range.
Connect MMT162 to a calibrated current/voltage meter in order to
measure either current or voltage depending on the selected output type.
To calibrate analog output, see section Calibrate Analog Output on page
35.