User's Guide ______________________________________________________________________
6. Set the communication parameters according to the
configured port settings of WMT700.
The default values for WMT700 are:
- Bits per second: 9600
- Data bits: 8
- Parity: None
- Stop bits: 1
- Flow control: None
7. Click Apply and OK.
8. On the Settings tab, click ASCII setup. Select ASCII
sending - Send line ends with line feed. Click OK and
close the New Connection Properties window.
9. On the View menu, click Font. In the Font list, select
Terminal.
10. From the Call menu, click Call. Enter a name and select an
icon for the connection. Click OK.
When a serial line cable is connected to the PC and the terminal
setup is correct, switch the sensor power supply on. The following
information is displayed to terminal screen:
WMT700 v<version number>
The sensor goes to measurement mode after four seconds and
it is ready to receive measurement mode commands.
At the end of each command, press ENTER to execute the
command. For a successful execution, the following combinations
of Carriage Return <CR> and Line Feed <LF> are accepted:
- <CR>
- <LF>
- <CR><LF>
Communication baud rate can be configured from 300 baud to
115200 baud. For available baud rates, see Appendix D,
Configuration Parameters, on page 205.
108 __________________________________________________________________ M211095EN-E