Chapter 5 _________________________________________________________________ Operation
Entering and Exiting Configuration Mode
With OPEN and CLOSE commands you can toggle between
configuration mode and measurement mode. In the measurement
mode, OPEN changes the mode to the configuration mode, and in
the configuration mode, CLOSE changes the mode to the
measurement mode. Note that the configuration mode does not
recognize the OPEN command and the measurement mode does
not recognize the CLOSE command.
OPEN — Entering Configuration
Mode
To switch WMT700 from the measurement mode to the
configuration mode, use the CLOSE command.
$0OPEN<enter>
|| | |____
To activate the command, press Enter
|| |_________The OPEN command
||____________
Sensor address, all sensors answer 0 address
|_____________Fixed $ character starts command
After a successful OPEN command the sensor switches from
measurement mode to configuration mode and the following
symbol is displayed:
>
CLOSE — Exiting Configuration
Mode
To switch WMT700 from the configuration mode to the
measurement mode, use the CLOSE command.
>CLOSE<enter>
| | |____
To activate the command, press Enter
| |__________
The CLOSE command
|_____________
Configuration mode prompt
VAISALA _______________________________________________________________________ 109