User's Guide _______________________________________________________________________
32 ___________________________________________________________________ M210619EN-E
0505-276
Figure 13 Tightening the Clasp Nut
The following numbers refer to Figure 13 above:
1
=
Probe
2
=
Clasp nut
3
=
Fitting screw
4
=
Pen (for marking)
NOTE
Take care not to overtighten the clasp nut to avoid difficulties when
opening it.
CAUTION
Take care not to damage the probe body. A damaged body makes the
probe less tight and may prevent it from going through the clasp nut.
CAUTION
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.
NOTE
When HMT318 is installed in a process with pressure differing from
normal atmospheric pressure, enter the pressure value of the process (in
hPa or mbar) into the transmitter memory via the serial line (see section
Set Ambient Pressure for Calculations on page 51).