Vaisala HUMICAP HMT310 Guide de l'utilisateur Page 72

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 105
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 71
User's Guide _______________________________________________________________________
72 ___________________________________________________________________ M210619EN-E
Set Calibration Date
Syntax: CDATE yyyy mm dd<cr>
1. Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
74).
2. Issue the LI command and set the offset value to 0 and gain value
to 1.
3. Issue the CDATE command and set the calibration date.
Example:
>cdate 2001 12 11
Calibration : 2001-12-11
>
Analog Output Calibration
Syntax: ACAL<cr>
Connect the HMT310 to a multimeter in order to measure either current
or voltage depending on the selected output type. Issue the ACAL
command. Then type the multimeter readings and press Enter.
Example (current outputs):
>acal
Ch1 I1 (mA) ? 2.046
Ch1 I2 (mA) ? 18.087
Ch2 I1 (mA) ? 2.036
Ch2 I2 (mA) ? 18.071
>
Vue de la page 71
1 2 ... 67 68 69 70 71 72 73 74 75 76 77 ... 104 105

Commentaires sur ces manuels

Pas de commentaire