HMP230 SERIES
Appendix 4: Digital current loop module M210225en-B
118
1. INSTALLATION
Switch the transmitter off.
Resistors R2, R3 and R4 between connectors X4 and X5 in the component
board in the cover of the transmitter are removed with side-cutting pliers. The
module is plugged in connectors X4 and X5 on the main board of the HMP230
transmitter; connector X1 on the module board to connector X4 and connector
X2 to connector X5.
RX GND TX
R2
R3
R4
X6
RX +
TX +
X4
X5
Current loop
module
X1
X2
DATA LOOPS
TX -
RX -
X17
Connect the data wires to screw terminal X6 on the main board. Switch the
power on.